9. Troubleshooting

Table 9 ·

Trouble Cause Solution Note
Current signal noise Air bubble adsorbs on the rotating electrode surface. · Take out the electrode from the solution, rotate the electrode over 1000 rpm, and then dip the rotating electrode into the solution while rotation. · Air bubbles on the electrode surface are formed easily after the gas purge.

· Air bubbles may occur during the electrochemical measurements sometimes.
Volume of solution is not suitable. · 65 ± ml solution is suggested to use for avoiding the data noise, when the cell (100 ml) in the RRDE-3A Ver. 1.2 package is used. · Dipping around 5 mm of the rotating electrode tip into the solution is supposed to avoid the turbulence noise effectivelly.
Dirty or unevenness of electrode surface. · Polish the electrode.

· If the electrode surface is modified, please re-modify the electrode so that the modification layer is thin and flat.

Rotating electrode is attached to the shaft in askew direction. · Remove the electrode from the shaft, and connect it to the shaft again.
RRDE-3A Ver. 1.2. is not well grounded. · Connect the RRDE-3A Ver. 1.2. and potentiostat to the same ground.
Air bubble adsorbs on the reference electrode tip. · Flip the reference electrode and move off the air bubble. · Alteration of the sample solution temperature may cause generation of the air bubbles.
Defective reference electrode. · Exchange the reference electrode for a new one.
The connecting resistance between carbon brush and shaft is changed after long time use. · Remove the carbon brushes from the shaft, clean the carbon powder on the shaft, and then setup the carbon brushes up again.

· In the case that carbon brushes are consumed greatly by frictionafter long time use, please replace them.
· The damage of the carbon brushes may possibly be induced by the big shock during the transportation.
Shaved carbon brush powder is piled on the bearing surface after long time use. · Open the motor cover and remove the shaved carbon powder with such as a cotton swab. · The carbon powder stored on the bearing may possibly cause the ring current noise in most cases.
The big vibration of the shaft tip after long time use. · Adjust and fix the screw of the shaft to motor. If not help, please replace the shaft. · The shaft damage may possibly occur due to the big shock of the transportation.

· Distorted contact pin on the shaft often causes the vibration.
Adnormal lower current High resistance between carbon brushes and shaft. · Remove the carbon brushes, and then set new ones up.
Different quantity of redox agents. · Prepare the sample solution again.
High resistance contact pin on the shaft and disk electrode. · Exchange the shaft for a new one.
Current over flow The reference electrode is not immersed in the solution. · Dip the reference electrode tip into the electrolyte solution.

· If the solution volum is not enough, please increase the solution untill soaking the electrode tips.

Unsuitable potentiostat setting, or fault of the potentiostat. · Adjust the current sensitivity setting.

· Fix the potentiostat or update the application software of the potentiostat.

No current flow The rotating electrode or counter electrode is not immersed in the solution. · Dip the tip into the electrolyte.

· If there is shortage of the solution, please increase the solution untill soaking the electrode tips.

Disconnection of the electrodes with electrode leads of potentiostat. · Connect them securely.




Click here to contact ALS.