Standard Composition

A typical miniature spectrophotometer consistis of four basic elements:
– SEC2000 (detector)
– Visual Spectra 2.1 (program software)
– SEC2000-TH (light source for Visible)or SEC2000-DH (light source for UV-Visible
– SEC2000-CUV (cuvette holder)

Item Description Item Code
Detector SEC2000 Spectrometer SEC2000
Software Visual Spectra 2.1 Visual Spectra 2.1
Light Source Tungsten-Halogen Lamp (for Visible) SEC2000-TH
Deuterium & Tungsten Halogen Lamp (for UV-Visible) SEC2000-DH
Sample Holder 3-Way Holder SEC2000-CUV


Spectrometer

MODEL SEC2000-UV/VIS SEC2000-VIS/NIR
Detector 2048 pixels CCD array
Wave length range 200 - 900 nm 400 - 1000 nm
Grating Blazed at 400 nm Blazed at 500 nm
Resolution ∼ 2.3 nm: slit 50 µm x 1000 µm (standard) (*1)
Accuracy < 1% at 1 Abs
Dark noise < 2 mAbs
A/D resolution 14 bit
Optical entrance SMA905
Interface USB 2.0
Operating system WindowsTM 8.1 / 10
Dimension (W x D x H) 98 x 118 x 35 mm
Catalog No. 013256 013251

(*1) Slit could be selected from: 10, 25, 100, 200 µm.

Light source

For UV/VIS mode:

SEC2000-DH (*2)
Light source (deuterium & tungsten halogen)
Spectral range (Halogen + D2) 200 - 1100 nm
D2 lamp Spectral range 200 - 400 nm
Power consumption (240 nm) > 5 x 10-8 W/nmsr
Stability 1 x 10-3 AU
Drift < 0.25%/h
Bulb life (*3) > 800 h
Halogen lamp Bulb life > 2000 h
Dimension (W x D x H) 98 x 118 x 35 mm
Catalog No. 012193

(*2) Standard SEC2000-DH does not include fiber-optic connector. Selection of SMA905 type
SEC2000-DH is possible.
(*3) Value for less than 50% power consumption of 240 nm.

For VIS/NIR mode:

SEC2000-TH (*4)
Light source (tungsten halogen)
Spectral range 360 ~ 2000 nm
Bulb life 1,500 h
Connector SMA905
Dimension (W x D x H) 98 x 118 x 35 mm
Catalog No. 012194

(*4) SEC2000-TH includes fiber connector (SMA905).

Visual Spectra 2.1

Visual Spectra is the software for SEC2000. The minimum recommended PC configuration, for the
instalation of the Visual Spectra 2.1 is windowsOS 8.1, CPU Intel Core i3.

Calibration mode

SEC2000 Calibration mode

Using the calibration mode, you could have the concentration analysis for two ways. The concentration
of the sample could be obtained by baseline method or by the absorbance calue at a single wavelength.
If your measurement analysis has a baseline method, in this case the noise will not influence the result.
For non-distinct and unclear absorption peak, you could determine the absorbance at the sinle wavelength.

Real time Monitoring

SEC2000 Real time Monitoring

The oxidation of potassium Ferrocyanide at 800 mV vs Ag/AgCl was measured by absorbance mode using SEC2000.

last modified 2020/04/24