Reference electrode selection - Ag/AgCl type II


This is the standard type of aqueous reference electrode in Polymethyl pentene insulation and ceramic liquid junction.

013613 013488 013691
RE-1BP Reference electrode (Ag/AgCl)

RE-3VT Reference electrode screw type (Ag/AgCl) RE-1CP Reference electrode (Ag/AgCl/Saturated KCl)
RE-1B Reference electrode (Ag/AgCl)
RE-3VT Reference electrode screw type (Ag/AgCl)
RE-1CP Reference electrode (Ag/AgCl/Saturated KCl)
Reference potential
AgCl+e=Ag+Cl-
E0=196 mV vs RHE (25 deg C)
AgCl+e=Ag+Cl-
E0=196 mV vs RHE (25 deg C)
AgCl+e=Ag+Cl-
E0=198 mV vs RHE (25 deg C)
The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597).}
-Material: Polymethylpentene is used, because of the excellent chemical resistance characteristics.
-Liquid junction: Ceramics is used. (∗If you keep the electrode immersed in an alkaline solution for a long time, life time will be shortened.)
Aqueous reference electrode with a 3 M NaCl as an internal solution.


Aqueous reference electrode specific for ALS line-up flow cell, as Cross flow cell, Radial flow cell, EQCM flow cell and SEC-Spectroelectrochemical flow cell. Aqueous reference electrode with a saturated KCl as an internal solution.