Application - Synchronizing EC and Spectrum

SEC2020 Spectrometer system and Model 2325 Bi-Potentiostat
Synchronized

Cyclic voltammetry

and

Spectrum

Measurement


Trigger connection

The SEC2021 Spectrometer supports Trigger Mode for synchronized measurement with external device. It is a function to synchronize spectroscopic measurement and tie measurement with an external measurement.

Pin No. Signal Definition Function
1 Power 5V Output When connecting to the PC's USB port, this pin is connected to VBUS. Up to 100 mA current
2 No connection
3 No connection
4 Output Shutter Shutter ON/OFF
5 Output D2_ON D2 light source ON/OFF
6 Output HAL_ON Halogen light source ON/OFF
7 Input Trigger-IN TTL
8 GND GND Ground

SEC2021 Spectrometer Port Pin number

Fig. 1 SEC2021 Spectrometer Port Pin number.


Measurement example using cuvette type spectroelectrochemical cell

UV-visible absorption spectrum and absorption change on electrode reaction can be obtained by using optically transparent electrode(OTE). Gold or Platinum mesh electrode was used as an OTE. Cyclic voltammetry and Absorbance of the 2 mM potassium ferricyanide, as the reference of the absorbance, performed in a SEC-C Thin Layer Quartz Glass Spectroelectrochemical cell are shown below.

Electrochemistry Spectrometry
Cyclic voltammetry for 2 mM potassium ferricyanide.
Fig. 1 Cyclic voltammetry for 2 mM potassium ferricyanide.
Absorption spectra of the electrolytic balance for 2 mM potassium ferricyanide electrolyzed at different potential.
Fig. 2 Absorption spectra of the electrolytic balance for 2 mM potassium ferricyanide electrolyzed at different potential.
Sample
Test solution : 2 mM Potassium ferricyanide / 1 M KNO3
Measurement parameter
Electrodes
Working electrode: SEC-C Pt Gauze working electrode
Reference electrode: RE-2BP Calomel Reference electrode
Counter electrode: SEC-C Pt counter electrode

Electrochemical parameters
Technique: Cyclic voltammetry
Initial E : +500 mV
Final E : 0 mV
Scan Rate (Volt) : 5 mV/S
Sensitivity : 10-4
Setting
Measurement: Spectrum
Setting: Trigger (SW Level trigger)
Time sequence: Sampling interval: 0.02 s
                          Measurement time: 600 s








Synchronized measurement

In this section connection of SEC2020 Spectrometer system with Potentiostat for a synchronized electrochemistry and spectrochemistry measurement is introduced and explained.
The spectrometer supports Trigger Mode for synchronized measurement with external device. Depending on the trigger mode, you can control acquisition of spectrometer data with external Input/Output signal. It makes possible to control the spectrometer measurement with external operation of the external devices.

As an example Interface 1010ETM (Gamry Instruments), Model 2325 Bi-Potentiostat (ALS Co., Ltd.) and Model 760D Bipotentiostat were used.
For detailed explanation, choose one of the potentiostat below: